Description
Areas of application
- Standard instrumentation for pressure measurement: Monitoring of steady pressures and dynamic processes (e.g. stall detection or blade profile pressures) on all our test rigs.
Functionality
-
Resistive pressure measurement: The stretching and compression of the pressure diaphragm causes a change in resistance, which is translated into a voltage signal
-
Piezoresistive pressure measurement: The deformation of the diaphragm also causes a change in electrical conductivity, which results in a greater change in resistance and increases the measurement resolution
Properties
-
Precise integration: Sensors are integrated and positioned by our in-house sensor workshop without influencing the geometry around them
-
Modular design: Steady pressure transducers can be flexibly adapted to different test rigs
-
Compact miniature sensors: High measuring accuracy for dynamic flow analyses
Specification
Category | Steady | Unsteady |
---|---|---|
Existing measurement ranges | 10“ WC, 20“ WC, 1 PSI, 2,5 PSI, 5 PSI, 10 PSI, 15 PSI, 30 PSI, 45 PSI, 100 PSI | 5 PSI, 10 PSI, 25 PSI, 50 PSI |
Resolution | 0,003 % FS | infinitesimal |
Measurement accuracy | 0,15 % FS ≤ 2,5 PSI 0,05 % FS > 2,5 PSI | 0,1 % FS |
Sampling rate | max. 100 measurements/channel/second | max. 100 measurements/channel/second |