Description
Areas of application
-
Pressure measurement in flow fields: Determination of total, static and dynamic pressure
-
Flow vector measurement: recording the speed and direction of the flow
Functionality
Measurement variables: Total pressure, static and dynamic pressure, velocity vectors, total temperature (for probes with thermocouple)
- Measurement principle:
- Total pressure via a tube directed perpendicular to the flow
- Static pressure from wall pressure holes
- Dynamic pressure from the difference between the two values
-
Total temperature via integrated thermocouple
- Calibration:
- High-speed calibration channel (HGK) of the Institute for Precision Tuning
- Numerical calibration for flow probes
Properties
Adjustable: In-house developed adjustment mechanisms for radial and peripheral probe adjustment
Flexible use: Adaptable to different flow conditions and test bench configurations